Abstract
In preparation for the 2017 total solar eclipse, St. Catherine University developed a calibration protocol for the temperature sensors flown during thermal wake boom experiments. The calibration method used a standard two-point technique that corrected each individual sensor for both slope and offset errors using a high quality NIST certified thermocouple as the temperature standard. Our method is not absolute but corrects each sensor relative to the NIST standard so that we feel some confidence that individual sensor variations are mitigated. In preparation for the eclipse, calibration curves were generated for over 200 individual digital and thermistor temperature sensors.
Previous Versions
Calibration of Temperature Sensors in Preparation for the 2017 Total Solar Eclipse
In preparation for the 2017 total solar eclipse, St. Catherine University developed a calibration protocol for the temperature sensors flown during thermal wake boom experiments. The calibration method used a standard two-point technique that corrected each individual sensor for both slope and offset errors using a high quality NIST certified thermocouple as the temperature standard. Our method is not absolute but corrects each sensor relative to the NIST standard so that we feel some confidence that individual sensor variations are mitigated. In preparation for the eclipse, calibration curves were generated for over 200 individual digital and thermistor temperature sensors.